Abstract

Mucous samples collected through nasopharyngeal (NP) swabs are considered gold standard specimens for the detection of respiratory pathogens. Matrices of these highly viscous samples often cause significant background noises in immunoassays, especially immunoassays with high sensitivity. We demonstrated such nonspecific background signals in both a chemiluminescence enzyme-linked immunosorbent assay (ELISA) and a novel highly sensitive immunoassay called Microbubbling SARS-CoV-2 Antigen Assay (MSAA). We developed and demonstrated the effectiveness of two quick sample pretreatment methods, filtration and preadsorption, to decrease nonspecific signals and increase the signal-to-noise ratio (SNR). Using these pretreatment methods, the SNR (at 3.6 × 104 copies/mL of inactivated SARS-CoV-2) was increased by 42.4-fold (95% CI 41.0-43.8) and 67.1-fold (95% CI 57.9-76.3) in the MSAA, and 1.3-fold (95% CI 0.9-1.7) and 1.8-fold (95% CI 1.6-2.0) in the chemiluminescence ELISA assay. Sample pretreatment methods developed in this study are broadly adaptable for the development of immunoassays for highly viscous samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.