The Cepheid Xpert® Xpress SARS-CoV-2 assay is 1 of the several real-time reverse transcription polymerase chain reaction (RT-PCR) assays that received Emergency Use Authorization from the United States Food and Drug Administration (FDA) for detection of SARS-CoV-2. Here we report 4 SARS-CoV-2 samples that were reported as presumptive positives on the Cepheid platform while reported as positives on alternative RT-PCR platforms. Whole genome sequencing indicated that the samples were Delta variants and had point mutations in the N gene which potentially interfered with SARS-CoV-2 detection. Two types of point mutations were found in these samples in the US CDC 2019-nCoV Real time PCR N2 Probe region: C29203T and C29200T. C29203T is a novel point mutation, and C29200T has not been previously reported in the Delta variants. This underlines the fact that mutations in the real-time RT-PCR assay target region could hinder accurate detection of SARS-CoV-2.