Abstract
Hyperinflation of the laryngeal mask airway cuff may exert high pressure on pharyngeal and laryngeal structures. In vitro data show that high intra cuff pressures may occur when inflated to only 30% of the manufacturer's recommended maximum inflation volume. We prospectively assessed the pressure volume curves of paediatric sized laryngeal mask airways (size 1-3) in 240 consecutive children (0-15 years). Following laryngeal mask airway insertion the cuff was inflated with 1-ml increments of air up to the maximum recommended by the manufacturer. After each ml cuff pressure was measured. At the end all cuff pressures were adjusted to 55 cmH(2)O. The maximum recommended volume resulted in high intracuff pressures in all laryngeal mask airway brands and sizes studied. Approximately half the maximum volume produced a cuff pressure > or = 60 cmH(2)O. This occurred in all brands and all sizes studied. We recommend that cuff manometers should be used to guide inflation in paediatric laryngeal mask airways.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.