Abstract

A pressure-limited ventilator (Bourns BP-200, Anaheim, CA) and a simple, manually operated constant flow ventilator were studied using a commercially available infant lung simulator (Bourns LS-130, Anaheim, CA). The characteristics of inspiratory pressure-time relationships during ventilation with these ventilators were analyzed. The analytic properties of the inspiratory phase of the pressure-time curve were related to simulator resistance and compliance. A mathematical model was constructed relating the initial rapid rise of the pressure curve to simulator resistance and the subsequent slope of the curve to simulator compliance. Qualitatively similar tracings were obtained with the BP-200 during normal ventilation, with simulated airway obstruction and thoracic restriction. These findings suggest that monitoring pressure-time relationships may be useful in the qualitative assessment of resistance and compliance during pressure-limited ventilation of neonates.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.