Abstract

Cardiac output (CO) can be measured using the pressure recording analytical method (PRAM), which is a new, less invasive technique allowing beat-by-beat stroke volume monitoring from the pressure signals recorded in femoral or radial arteries. We investigated PRAM by comparing its cardiac output (PRAM-CO) with paired measurements obtained by electromagnetic flowmetry (EM-CO) and by standard thermodilution (ThD-CO) during various haemodynamic states in a swine model. Nine pigs were monitored with a pulmonary artery catheter and a femoral artery catheter at baseline, in a hyperdynamic state produced by administration of dobutamine and in a hypodynamic state induced by progressive exsanguination. Bland-Altman analysis was used. One hundred and eight paired cardiac output values over a range of EM-CO of 1.8-10.4 litre min(-1) resulted. We found close agreement between the techniques. Mean bias between EM-CO and PRAM-CO was -0.03 litre min(-1) (precision 0.58 litre min(-1)). The 95% limits of agreement were -0.61 to +0.55 litre min(-1). Similar results between ThD-CO and PRAM-CO were found. In a porcine model we have demonstrated accuracy of PRAM during various haemodynamic states. PRAM is a reliable tool to detect changes in cardiac output in pigs and has ability as a basic research tool.