Abstract

High flow nasal cannula therapy has a mild airway pressurization effect. These relatively low pressures can compromise the effectiveness of the technique, especially in those clinical scenarios where greater airway pressurization is required. Finding ways to increase this pressure would allow its use in patients with high pressurization requirements. We have therefore analyzed the increase in pressure achieved by coupling Boussignac® CPAP to the high-flow device circuit. Our preliminary results have shown that the pressure measured at the nasal cannula achieved with the AIRVO 2® can be significantly increased by associating Boussignac® CPAP. The FiO2 was stable and maintained during the study. We did not observed water condensation into the tubes. Studies are needed to confirm these results and their involvement in patients with various types of acute respiratory failure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.