This paper describes a novel method that applies pressure-assisted field-amplified sample injection with reverse migrating micelles (PA-FASI-RMM) for the online concentration of neutral analytes in MEKC with a low-pH BGE. After injection of a plug of water into the separation capillary, negative voltage and positive pressure were simultaneously applied to initialize PA-FASI-RMM injection. The hydrodynamic flow generated by the positive pressure compensated the reverse EOF in the water plug and allowed the water plug to remain in the capillary during FASI with reverse migrating micelles (FASI-RMM) to obtain a much longer injection time than usual, which improved stacking efficiency greatly. Equations describing this injection mode were introduced and were supported by experimental results. For a 450-s online PA-FASI-RMM injection, three orders of magnitude sample enhancement in terms of peak area could be observed for the steroids and an achievement of detection limits was between 1 and 10 ng/mL.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.