Abstract

To report pressure and stiffness, in healthy volunteers, of a new compression device with an air bladder inflated by a pump to regulate pressure. The device was applied to 60 legs of 30 volunteers and set to exert different pressures of 20-50mmHg. The exerted pressure was measured in supine and standing positions and during simple physical exercises; static stiffness index, dynamic stiffness index, and walking pressure amplitudes were calculated. The exerted pressure showed a good correlation with the expected pressure at each pressure range. The stiffness indices were >10mmHg in the range of inelastic materials. The device was considered very easy to apply and use by the testing researchers. The device stiffness is in the same range as the inelastic bandages. Consequently, similar hemodynamic effectiveness could be expected but must be proved. Unlike inelastic bandages, this device was easy to apply and use.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.