Abstract

A 58-year-old man had a bipolar epicardial left ventricular (LV) implanted for cardiac resynchronization therapy after a failed transvenous approach. The system was programmed in an LV tip-right ventricular coil configuration, but during follow-up loss of LV capture with recurrence of symptoms occurred. Changing to an LV tip-LV ring configuration restored LV pacing through anodal capture. Loss of cathodal capture was caused by dislodgment of the cathodal electrode due to a broken fixation suture. Anodal capture was used to reinstall resynchronization therapy, which resulted in clinical improvement. There were no adverse effects from anodal stimulation in a follow-up period of 6 months.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.