Abstract

At the PTB radiometric scales are compared which are based on electrodynamics (electron storage ring BESSY), on thermodynamics (black body of known temperature T90 of the International Temperature Scale) and on a thermal detector in combination with the electrical substitution principle (cryogenic radiometer). The uncertainty contributions of the different input parameters for establishing the independent radiometric scales are analysed in detail. According to this analysis spectral radiant powers in the visible are realized or expected to be realized (black body) with uncertainties (1 σ level) of 0,10% (BESSY), of 0,07% (black body, λ>650 nm) and of 0,007% (cryogenic radiometer). Filter radiometers calibrated at the three independent primary standards are used as transfer instruments for the comparison. The measured values of the responsivities of the filter radiometers are estimated to be uncertain by 0,11% (BESSY), 0,08% (black body, λ>650 nm), and 0,03% (cryogenic radiometer). Preliminary experimental results of the comparison between BESSY and the cryogenic radiometer are in agreement with the uncertainty analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.