Abstract

We report on the eighth cryogenic run of MiniGRAIL, which is planned to be the first science run. Three new capacitive transducers were mounted on the sphere, all coupled to superconducting transformers and a double-stage SQUID amplifier. All SQUID modules had a noise level between 1 and 2 μΦ0 Hz−1/2 at 4 K in separate test runs. The detector was upgraded with a complete data acquisition and data storage system, and a data analysis program for millisecond burst signals was written using a matched filter. Two cosmic-ray detectors were built and installed on the roof above MiniGRAIL as a veto for cosmic rays. A test run was done, but failed because of bad electrical isolation of the transducer electrode at low temperatures. A new cool-down is currently going on. With an already obtained sphere temperature of 65 mK and a SQUID noise of 100ℏ, the expected sensitivity should be 4 × 10−22 Hz−1/2.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.