Abstract

Nearly transparent cellulose aqueous solution was prepared using the clear layer of diluted cellulose dispersion solution produced by wet jet milling. Glucose oxidase (GOx) was immobilized on platinum-iridium electrode by applying certain amount of cellulose aqueous solution containing GOx and dried for at least two days in room temperature. The obtained GOx-immobilized electrodes provided good response current up to 22.4 mM. Good linear relationship between glucose concentration and response current was observed. Correlation coefficient of 0.994 was obtained ranging from 0 to 11 mM on the electrode prepared using 500 mg/L cellulose and 250 mg/L GOx mixture solution. The sensor response was significantly fast and steady state current was obtained within one second. The stability of GOx-immobilized electrode response was measured for 30 days. After initial decrease of response current for the first few days, relatively constant responses were obtained for twenty days.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.