Abstract

The aim of this study was to investigate the suitability of picosecond lasers for hard-tissue preparation. Using scanning electron microscopy (SEM) and histological studies, the quality of ablation was examined. There was no thermal or mechanical damage observed. The ablation rates of enamel, dentine and carious enamel and the temperature rise during tissue removal were measured. Using a fast silicon photodiode, the lifetime of the plasma was determined to be 11±3 ns (Nd-YLF, 1.25 mJ pulse−1, pulse duration 30 ps). In order to achieve a better understanding of the ablation process, plasma parameters such as electron density [(5±3)×1017 cm−3] and excitation temperature (4500±2000 ℐg, Nd-YLF, 0.3 mJ pulse−1) were obtained from emission spectra. By examining the line width of singly ionized calcium lines, a difference was found in the ablation of healthy or artificially demineralized enamel, thus allowing the selective removal of caries.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.