Abstract
The technique which is described combines the advantages of the techniques formerly proposed in the literature in each stage of the preparation of transmission electron microscopy (TEM) specimens including a single metal-ceramic interface. It allows easy handling of the thin foils in spite of their brittleness. Preferential thinning of the softer material in the two-phase foil is prevented, and both sides of the interface are thinned down to comparable thicknesses. The nickel-alumina bicrystal interface observed in TEM is neat and free from any reaction layer. This method is easily adaptable to other metal-ceramic systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.