Abstract

An advanced electrochemical sensor for the detection of enrofloxacin (ENR) based on the use of a modified electrode containing cadmium sulfide (CdS) nanoparticles (NPs) is reported. The CdS NPs were synthesized and characterized and then coated onto the electrode to fabricate a modified electrode that exhibited a lower limit of detection of 9.5×10-8 mol·L-1 . This detection limit compares with a traditional electrode that exhibited a concentration detection range of 1.0×10-2 to 1.0×10-7 mol·L-1 . This modified electrode demonstrated good selectivity, reproducibility, response time (<40s), lifetime (up to 12wk), and pH range (3.3-7.2) for the determination of ENR in real samples (eg, pig urine).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call