Abstract

A novel diclofenac ion-selective electrode is prepared, characterized and used in pharmaceutical analysis. The diclofenac complex with hexadecylpyridinium bromide is obtained in situ by soaking the PVC-membranes in a 1 × 10 −2 M diclofenac solution. Among four different solvent mediators tested, dibutyl phthalate (DBP) exhibited a proper behavior including Nernstian slope of the calibration curve, fast response time and good reproducibility of the emf values. The electrode exhibits a Nernstian slope of −59 ± 1 mV decade −1 for diclofenac in the concentration range 1.0 × 10 −5 to 1.0 × 10 −2 M with a limit of detection of 4.0 × 10 −6 M. The electrode displays a good selectivity for diclofenac with respect to a number of common inorganic and organic species. It can be used in a pH range of 6.0–9.0. The membrane sensor was successfully applied to the determination of diclofenac in its tablets as well as for its recovery from blood serum and urine samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.