Abstract

A standard material containing chrysotile asbestos for the validation of x-ray diffractometric quantitation was developed using an asbestos-containing building material i.e., perlite board. The board as the base material was crushed, pulverized, and homogenized. The homogeneity of the powder of perlite board was estimated by analysis of variance. The diffraction intensity values of the crystalline phases and the concentrations of elements determined by x-ray diffractometry and x-ray fluorescence analysis were used for analysis of variance. There is no significant difference between the within-bottle variance and the between-bottle variance, indicating that the powdered perlite board was sufficiently homogenous. The concentration of chrysotile in the material was determined using two methods: an internal standard/x-ray diffractometry method and the x-ray diffractometry/Rietveld refinement. The concentration of chrysotile in the material was determined by an internal standard/x-ray diffractometry method and the material had a chrysotile concentration of 24.1 ± 0.2 mass%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.