Abstract

A home-made column (3.2 mm i.d., 100 mm length) was prepared in which glycidyl polymethacrylate was used. The optimum value of the initiator corresponded to 5 mg/mL. The column morphology was characterized by scanning electron microscopy. The permeability was evaluated using acetonitrile and water as a mobile phase and uracil as an unretained substrate. A simple and economical reversed-phase HPLC method has been developed for the simultaneous estimation of paracetamol and chlorzoxazone in their pharmaceutical formulations. Components were determined using a UV detector at 270 nm. The mobile phase was composed of 1% formic acid solution and acetonitrile (65/35, v/v); 0.7 mL/min flow rate and 5.0 µL injection volume were used. Peak resolution was 1.96. All findings allow concluding that the prepared stainless steel conventional HPLC column and the novel validated method are applicable for quality control and routine analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.