Abstract

We performed prenatal diagnosis of organic acid disorders using two mass spectrometric methods; gas chromatography mass spectrometry (GC/MS) and tandem mass spectrometry (ESI/MS/MS). Of 28 cases whose amniotic fluid was tested, 11 cases were diagnosed as “affected”. All cases whose samples were diagnosed as “unaffected” were confirmed to have no symptoms or abnormalities in urinary organic acid analysis after birth. Of the 11 “affected” cases, two cases were missed by ESI/MS/MS but not by GC/MS. When the stability of metabolites in amniotic fluid was checked, it was found that acylcarnitines degraded in one week at room temperature, whereas organic acids such as methylmalonate or methylcitrate were stable for at least 14 days. Prenatal diagnosis by analysis using simultaneous two or more methods may be more reliable, though attention should be paid to sample transportation conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.