Abstract

A 63-year-old woman with a St. Jude Medical Riata 1570 right ventricular lead complained of intermittent hiccups 2 months after implant. Interrogation revealed elevated pacing threshold and diaphragmatic stimulation. Pacing and shock lead impedances remained stable. No inappropriate sensing was noted. Fluoroscopic examination of the lead revealed a thin radio-opaque wire seen between the 2 defibrillator coils away from the main body of the lead. After extraction, a tear in the insulation of the lead was noted allowing the inner wire to protrude. This case illustrates a novel mechanism of insulation failure without inappropriate sensing or impedance change.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.