Abstract

Liquid argon (LAr) is an attractive target for the direct detection of WIMPs. A LAr prototype detector was designed to study the technology and properties of LAr detectors. The prototype detector had an active volume containing 0.65 kg of liquid argon. A liquid nitrogen (LN) cooling system allowed the temperature fluctuation of the liquid argon to be controlled within less than 0.1 K during a one month run. In the 22Na calibration run, the LAr prototype obtained 1.59±0.02 p.e./keV light yield for 511 keV gamma rays using a domestic-made argon purification system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.