Abstract

A new method has been developed for the monitoring of glutathione S-tranferase (GST) detoxification activity toward styrene oxide (SO). The enzymatic reaction was carried out directly in a thermostatted autosampler vial and the formation of conjugates between glutathione (GSH) and SO was monitored by sequential MEKC runs. The determinations were performed in a 50-microm fused silica capillary using 50 mM SDS in 20 mM phosphate 20 mM tetraborate buffer (pH 8.3) as a background electrolyte; separation voltage 28 kV (positive polarity), temperature of capillary 25 degrees C, and detection at 200 nm. The method is rapid, amenable to automation, and requires only small amounts of samples, which is especially important in the case of GST isoenzyme analyses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.