Abstract

To survey the whole-body positron emission tomography/computed tomography (PET-CT) examinations in Jiangsu Province and establish the diagnostic reference levels (DRL). A total of nine hospitals in southern, central and northern Jiangsu Provinces were selected. Activity duration product (ADP), activity per body weight (AW), dose-length product (DLP) and total effective dose (ET) were choosen to establish DRLs for whole-body PET-CT examinations. DRL and achievable dose (AD) were calculated according to International Commission on Radiological Protection report. The conversion coefficient method was used to calculate the ET induced by PET-CT whole-body imaging. DRLs of whole-body PET-CT examinations in Jiangsu Province were set to be 922mGy·cm (DLP), 4453MBq.min (ADP), 5.22MBq.kg-1(AW), 19.8mSv (ET). The ET for whole-body PET-CT examinations is higher than other countries. The effective dose from radionuclides (EF) in female patients is lower than that in male patients. Effective doses from CT(ECT), ET and ECT/ET in female patients are higher than that in male patients (P<0.01). Provincial DRLs are established for four radiation metrics in whole-body PET-CT examinations. DRL of AW in whole-body PET-CT examinations is at the medium level, and DLP is higher than that reported in most literature, which has the potential to be further reduced.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.