Abstract

The performance of normoxic PAG (nPAG) gel dosimeter with respect to spatial integrity, temperature sensitivity and dose-rate dependency makes it more reliable than PAG gel dosimeter. As a 3D dose distribution measurement in dosimetry, nPAG gel dosimeter with THPC as anti-oxidant, named PAGAT was studied. 2.9% mass concentration of the formaldehyde solution was added into PAGAT to solidify the gel, named PAGAT-f. The melting point of PAGAT-f polymer gel was increased by adding formaldehyde and can be stored at room temperature after gel preparation without refrigeration. After irradiation with a 9 MV LINAC, magnetic resonance imaging (MRI) is employed as readout modality. It’s found that by adding formaldehyde, R2-dose response of PAGAT-f is similar to PAGAT but the dose response sensitivity of PAGAT-f is better than that of PAGAT at low doses. The results prove the feasibility and good temporal stability of PAGAT-f for dose distribution measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.