Abstract

Background: The main purpose of the present study was to assess the short term performance of a cementless femoral stem in total hip replacement. Methods: Cross-sectional observational study of a 48-patient cohort with Phenom? femoral stems implanted between June 1, 2014 and September 1, 2018, to determine clinical performance, stability, and radiographic osseointegration. Patients were followed-up from 13 to 76 months (mean: 44.5 months) and assessed using the Harris Hip Score-HHS, the Hip Disability and Osteoarthritis Outcome Score-HOOS and radiographs. Results: All stems were radiologically stable. Mean Harris Hip Score was 89.8 and the HOOS was 80.4. No statistical differences were observed among patients with different diagnoses. Conclusions: The short-term results revealed satisfactory clinical outcomes and radiological signs of implant stability in all cases. Using two functional scores was useful in detecting biases and a low to moderate agreement was found between the scores.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.