Abstract

A measurement system for precise realization of quantum Hall resistance standard with GaAs heterostructures is described. A comparison with a standard Klitzing resistor is shown and discussed. The preliminary results indicate that the value of RH,2 realized in Hall plateau (i=2) is very close to CIPM value. Thus the technique will offer a more reliable and precise calibration service through transfer standard. Further attempts to improve the measurements for precise determination of RH are underway.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.