Abstract

The first in vivo experiments in support of a new technique for delivering stereotaxic hyperthermia have been conducted at the Experimental Surgery Facility of the University of Virginia's Medical Center. We call this technique the "Video Tumor Fighter." In each of twelve trials a single, small permanent magnet or train of small permanent magnets was implanted on the brain surface of adult canine models. In three of the trials, this "seed" (typically 6-mm diameter X 6-mm long) was moved by magnetic manipulation to different locations within the brain. In two other trials, the seed moved along the interface between the brain and the inner vault of the skull. The noncontact magnetic manipulation was accomplished by coupling the permanently magnetized seed to the large dc magnetic field gradient created by a water-cooled coil surrounding the animal's head. The seed's motions were monitored with x-ray fluoroscopy; its rate of movement was found to be approximately 0.8 mm s-1. The forces required to produce these motions were on the order of 0.07 N. We document here the instrumentation used in these trials, describe the experimental procedures employed, and discuss the technical aspects of the results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.