Abstract
Our preliminary experience is presented in the use of the Berger neurobiopsy device for ultrasound localization and biopsy of intracranial lesions through a burr hole. The apparatus and technique are described, along with the results of its use in the first 49 patients. In these patients 43 tumours were biopsied, all except one successfully. Three abscesses were aspirated, two intraventricular shunt catheters were sited and in one patient the diagnosis of postradiation gliosis was confirmed and tumour excluded. The advantages and limitations of the technique are discussed. It is advocated as a simple and time-saving alternative to CT stereotactic biopsy in many cases.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.