Abstract

Boron neutron capture therapy (BNCT) is an experimental technique for the treatment of certain kinds of tumors. Research in BNCT is performed utilizing both thermal and epithermal neutron beams. Epithermal neutrons (0.4 eV-10 keV) penetrate more deeply into tissue and are thus used in non-superficial clinical applications such as the brain glioma. In the last few years, the fast reactor TAPIRO (ENEA-Casaccia Rome) has been employed as a neutron source for research into BNCT applications. Recently, an 'epithermal therapeutic column' has been designed and its construction has been completed. The Monte Carlo code MCNPX was employed to optimize the design of the column and to evaluate the dose profiles and the therapeutic parameters in the cranium of the anthropomorphic phantom ADAM. In the same context, some preliminary evaluations of the undesirable doses to the patient were performed with MCNPX. A hermaphrodite phantom derived from ADAM and EVA was employed to evaluate the energy deposition in some organs during a standard BNCT treatment. The total dose consists of the contributions from the primary neutron beam, the neutron interactions with boron and the neutron induced photons generated in the epithermal column structures and in the patient's tissues. The paper summarizes the computational procedure and provides a general dosimetric framework of the patient radiological protection aspects related to a BNCT treatment scenario at the TAPIRO reactor.