Background In magnetic resonance (MR) image–guided microwave thermocoagulation of liver tumors, the choice of the optimal puncture route is an important and time-consuming process. To assist this process, we have developed a motorized MR-compatible manipulator. Methods The manipulator consists of a passive end effecter with 2 degrees-of-freedom (DOF) rotation and active base stages with 3 ultrasonic motors. It automatically chases the preset target point with synergetic remote-center-of-motion (RCM) control. A mechanical torque limiter and an electrical shutdown switch were added for patient safety. Results The manipulator was used for this procedure in 15 cases and successfully utilized to treat liver tumors in various locations. Thoracoscopic assistance was combined with the manipulator in 6 cases. No complications were experienced. Conclusions The manipulator was found to be very effective for assisting MR-guided microwave coagulation of liver tumors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.