Abstract
The GORE EXCLUDER Iliac Branch Endoprosthesis (IBE) device is designed to seal off a common iliac artery (CIA) aneurysm, preserving the internal iliac artery during endovascular aortic repair. We report the case of an 84-year-old man with isolated saccular right CIA aneurysm (35 mm) and a relatively small terminal aorta (24 mm). The IBE device was successfully placed, and intraoperative angiography revealed no leakage or delay. However, postoperative computed tomography revealed marked compression of the contralateral leg by a bridging component. Although his ankle–brachial index was preserved, its acute occlusion was judged highly possible; we decided to perform preemptive angioplasty. The angiography revealed the stenosis only in the left anterior oblique view, and angioplasty was uneventfully performed. The leg was successfully patent at 1-year follow-up. When compression by IBE and bridging component in the terminal aorta is expected, caution should be preserved at intraoperative angiography following the device deployment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.