Abstract

Objective: To compare the diagnostic accuracy of various transcutaneous bilirubin (TcB) nomograms for predischarge screening. Methods: The paired total serum bilirubin (TSB) and TcB measurements collected in neonates ≥35 weeks and ≥2000 g birth weight were analyzed. BiliCare™ bilirubinometer was used for TcB measurement. We chose the following nomograms for the study: Bhutani nomogram, Maisel's nomogram, Agarwal nomogram, Thakkar nomogram, American Academy of Pediatrics (AAP) nomogram within 3 mg/dl of phototherapy cutoff, AAP nomogram >70% of phototherapy cutoff and if TcB value is above 13 mg/dl. The diagnostic accuracy of these nomograms for TcB was compared with TSB plotted in the Bhutani nomogram. Results: TcB showed a positive correlation with TSB (Pearson correlation coefficient = 0.783). Bhutani nomogram, Maisel's nomogram and AAP (using within 3 mg/dL cutoff) nomogram showed good sensitivity and low false-negative rate while avoiding blood draws in most neonates. Conclusion: Bhutani nomogram, Maisel's nomogram, and AAP (using within 3 mg/dL of phototherapy cutoff) nomograms have comparable diagnostic accuracy for predischarge bilirubin screening in neonates.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.