Abstract

OBJECTIVENocturnal hypoglycemia can cause seizures and is a major impediment to tight glycemic control, especially in young children with type 1 diabetes. We conducted an in-home randomized trial to assess the efficacy and safety of a continuous glucose monitor–based overnight predictive low-glucose suspend (PLGS) system.RESEARCH DESIGN AND METHODSIn two age-groups of children with type 1 diabetes (11–14 and 4–10 years of age), a 42-night trial for each child was conducted wherein each night was assigned randomly to either having the PLGS system active (intervention night) or inactive (control night). The primary outcome was percent time <70 mg/dL overnight.RESULTSMedian time at <70 mg/dL was reduced by 54% from 10.1% on control nights to 4.6% on intervention nights (P < 0.001) in 11–14-year-olds (n = 45) and by 50% from 6.2% to 3.1% (P < 0.001) in 4–10-year-olds (n = 36). Mean overnight glucose was lower on control versus intervention nights in both age-groups (144 ± 18 vs. 152 ± 19 mg/dL [P < 0.001] and 153 ± 14 vs. 160 ± 16 mg/dL [P = 0.004], respectively). Mean morning blood glucose was 159 ± 29 vs. 176 ± 28 mg/dL (P < 0.001) in the 11–14-year-olds and 154 ± 25 vs. 158 ± 22 mg/dL (P = 0.11) in the 4–10-year-olds, respectively. No differences were found between intervention and control in either age-group in morning blood ketosis.CONCLUSIONSIn 4–14-year-olds, use of a nocturnal PLGS system can substantially reduce overnight hypoglycemia without an increase in morning ketosis, although overnight mean glucose is slightly higher.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call