Abstract

This paper presents an architecture for generalized predictive control for an active prosthetic socket system, based on a cost function performance index measure for minimization of residual limb tissue injury. Finite element analysis of a transtibial residuum model donned with a total surface bearing socket was used to provide controller training data and biomechanical rationale for deep tissue injury risk assessment, by estimating the internal deformation state of the soft tissues and the residuum-socket interface loading under a range of prosthetic loading instances. The results demonstrate the concept of this approach for interface actuation modelled as translational spring and damper systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.