Abstract

In proton therapy, patient quality assurance (QA) requires measuring the beam range, spread-out Bragg peak (SOBP), and output factor. If these values can be predicted by using sampling measurements or previous QA data to find the correlation between beam setup parameters and measured data, efforts expended on patient QA can be reduced. Using sampling data, we predicted the range, SOBP, and output factor of the proton beam. To obtain sampling data, we measured the range, SOBP, and output factor for 14 data points at each of 24-beam range options, from 4–28 cm. Prediction conformity was evaluated by the difference between predicted and measured patient QA data. Results indicated that for 60% of patients, the values could be predicted within 3% of dose uncertainty.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.