Abstract
Intradialytic hypotension is the most common acute complication during conventional hemodialysis treatment. Prediction of such events is highly desirable in clinical routine for prevention. This paper presents a novel prediction method of acute symptomatic hypotension in which the photoplethysmographic signal is analyzed with respect to changes in amplitude, reflecting vasoconstriction, and cardiac output. The method is based on a statistical model in which the noise is assumed to have Laplacian amplitude distribution. The performance is evaluated on 11 hypotension-prone patients who underwent hemodialysis treatment, resulting in seven events with acute symptomatic hypotension and 17 without. The photoplethysmographic signal was continuously acquired during treatment as was information on blood pressure and oxygen saturation. Using leave-one-out cross validation, the proposed method predicted six out of seven hypotensive events, while producing 1 false prediction out of 17 possible. The performance was achieved when the prediction threshold was chosen to be in the range 57%-65% of the photoplethysmographic envelope at treatment onset.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.