Abstract

The accuracy of dichotic multiple frequency auditory steady state in predicting pure-tone thresholds at 0.5, 1, 2, and 4.0 kHz compared to an ABR protocol (click and tone burst at 0.5 kHz) were explored in a group of 25 hearing-impaired subjects across the degree and configuration spectrum. Mean steady state thresholds were within 14, 18, 15, and 14 dB of the pure tones at 0.5, 1, 2, and 4 kHz, compared to the tone-burst ABR at 0.5 kHz pure-tone difference of 24 dB, and a click-evoked pure-tone (2-4 kHz) difference of 9 dB. Recording time for the steady state protocol was 28 minutes (+/- 11) compared to 24 minutes (+/- 9) of the ABR protocol. Degree of loss had a significant effect on steady state; configuration of hearing loss had a limited effect. Mf ASSR predicted thresholds with relative accuracy although some configurations showed discrepancies for low-frequency estimates.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.