Abstract
Pedicle screw fixation is one of the most common procedures used in spinal fusion surgery. The screw loosening is a major concern, which may be caused by broken pedicles. In vitro pullout tests or insertion torque are the main approaches for assessing the stability of the screw; however, direct evidence was lacking for clinical human spines. Here, we aim to provide a model that can predict the pullout strengths of pedicle screws in various pedicle conditions from X-ray images. A weighted embedded bone volume (EBV) model is proposed for pullout strengths prediction by considering the bone heterogeneity and confinement of the screw. We showed that the pullout strength is proportional to the EBV for homogeneous bone and the weighted EBV for layered composite bone. The proposed weighted EBV model is validated with in vitro Sawbones® pullout experiments. The results show that the model has better accuracy than the simple EBV model, with a coefficient of determination of 0.94. The proposed weighted EBV model can help assess the stability of a pedicle screw in a broken pedicle by simply examining 2D X-ray images.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Mechanical Behavior of Biomedical Materials
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.