Abstract

Ultrasound brain therapy is currently limited by the strong phase and amplitude aberrations induced by the heterogeneities of the skull. However the development of aberration correction techniques has made it possible to correct the beam distortion induced by the skull and to produce a sharp focus in the brain. Moreover, using the density of the skull bone that can be obtained with high‐resolution CT scans, the corrections needed to produce this sharp focus can be calculated using ultrasound propagation models. We propose here a model for computing the temperature elevation in the skull during High Intensity Focused Ultrasound (HIFU) transcranial therapy. Based on CT scans, the wave propagation through the skull is computed with 3D finite differences wave propagation software. The acoustic simulation is combined with a 3D thermal diffusion code and the temperature elevation inside the skull is computed. Finally, the simulation is validated experimentally by measuring the temperature elevation in several locations of the skull.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.