Abstract

Using the Lotmar moiré fringe interferometer and the potential acuity meter, I predicted the postoperative Snellen visual acuities in 30 implanted eyes that required secondary capsulotomies. The Lotmar interferometer was within one line of the postoperative visual acuity in 29 eyes (96.6%); the potential acuity meter prediction was within one line in 10 eyes (33.3%). An earlier study defined seven clinical conditions that produced false positive results with the laser interferometer. The same results occurred with the Lotmar interferometer. When the potential acuity meter was used, six of the seven conditions produced false positive results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.