Abstract

This work presents a solid phase extraction (SPE) method for simultaneous preconcentration of trace elements in water samples prior to their ICP-OES determination. Dowex 50W-x8 and Chelex-100 resins were used as SPE sorbent materials for preconcentration of trace Cd, Co, Cr, Cu, Fe, Ni, Pb and Zn. The optimum sample pH, eluent concentration and sample flow rates were found to 6, 3.0molL−1 and 3.0mLmin−1, respectively. In terms of multi-element preconcentration capabilities, Dowex 50W-x8 appeared to be a better sorbent. The recoveries for all the tested analytes were >95%. However, Chelex-100 showed a better performance in terms of recovery (>95%) towards Cu, Fe and Zn. Under optimized conditions using Dowex 50W-x8, the relative standard deviations for different metals were <3%. The limits of detection and limits of quantification ranged from 0.01–0.39μgL−1 and 0.05–0.1.3μgL−1, respectively. The accuracy of the preconcentration method was confirmed by spike recovery test and the analysis of certified reference materials. The SPE method was applied for preconcentration of the analyte ions in tap water, bottled water and wastewater samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.