Abstract

A rapid, sensitive and accurate atomic absorption method for Cr(III) and Cr(VI) ions was developed based on the cloud-point extraction (CPE) technique. Cr(III) reacts with a new Schiff's base ligand to form the hydrophobic complex, which is subsequently entrapped in the surfactant micelles. The Cr(VI) assay is based on its reduction to Cr(III) by the addition of concentrated H2SO4 and ethanol to the sample solution. The condensed surfactant phase containing the metal chelates is introduced into an atomic absorption spectrometer. The relative standard deviations were 2.1% for both species and the limits of detection were around 0.17 microg l(-1).