Abstract

Electrochemically activated glassy carbon electrodes adsorbed with alizarin red S (ARS) were applied to the electroanalysis of metal species. Copper species were readily extracted onto the ARS modified electrode and were determined electrochemically. Experimental results showed that the voltammetric peak current resulted from copper incorporation on ARS modified electrodes was significantly larger than that obtained on bare activated electrodes. Voltammetric peak currents showed linear response for copper concentrations in the range between 0.001 and 0.15 mM with a slope of 1.77 A M−1 (correlation coefficient: 0.998). The reproducibility of the voltammetric measurements was found to be within 10 % RSD for five replicate measurements on copper concentration of 0.10 mM. The modified electrodes could be applied for repeated copper determinations without appreciable deterioration in electrode response.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.