Abstract

A simple and sensitive preconcentration analysis-atomic absorption spectrometric procedure is described for the determination of lead, cadmium and nickel. The method is based upon on-line preconcentration of metal ions on a minicolumn of Cibacron Blue F3-GA immobilized on poly(hydroxyethylmethacrylate), poly(HEMA). The enrichment factors obtained were 42 for lead, 52 for cadmium and 63 for nickel (sample volume 10 mL and sample flow rate 5 mL/min). The relative standard deviations (n = 10), in 10 mL sample solutions containing 100 microg/L Pb(2+), 10 microg/L Cd(2+) and 100 microg/L Ni(2+) were 8.9, 3.7 and 3.5%, respectively. The limits of detection (blank + 3s) (n = 10), were found to be 12.01 microg/L for Pb(2+), 1.34 microg/L for Cd(2+) and 28.73 microg/L for Ni(2+). The accuracy of the system was checked with certified and tap water samples spiked with known amounts of metal ions. No significant difference was found between the achieved results and the certified values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.