Abstract

In intracardiac echocardiography (ICE) it may be beneficial to generate ultrasound images acquired at multiple frequencies, having the possibility of high penetration or high-resolution imaging in a single device. The objective of the presented work is to test two frequency-tunable probe prototypes in a preclinical setting: a rigid probe having a diameter of 11 mm and a new flexible and steerable 12-Fr ICE catheter. Both probes feature a forward-looking 32-element capacitive micromachined ultrasonic transducer array (aperture of 2 × 2 mm2) operated in collapse mode, which allows for frequency tuning in the 6-MHz–18-MHz range. The rigid probe prototype is tested ex vivo in a passive heart platform. Images of an aortic valve acquired in high-penetration (6 MHz), generic (12 MHz) and high-resolution (18 MHz) mode combine satisfying image quality and penetration depth between 2.5 cm and 10 cm. The ICE catheter prototype is tested in vivo using a porcine animal model. Images of an aortic valve are acquired in the 3 imaging modes with the ICE catheter placed in an ascending aorta at multiple depths. It was found that the combination of the forward-looking design and frequency-tuning capability allows visualizing intracardiac structures of various sizes at different distances relative to the catheter tip, providing both wide overviews and detailed close-ups.