Abstract

A liquid chromatography-tandem mass spectrometry (LC-MS/MS) has been developed and validated to determine the concentration of armillarisin succinate ester in mouse plasma and tissues, used for preclinical evaluation. Bavachin was employed as the internal standard. Separation was performed on a 3.5 µm Zorbax SB-C(18) column (30 × 2.1 mm), with a mobile phase consisting of methanol and aqueous 20 mm ammonium acetate. Both analyte and internal standard were determined using electrospray ionization and the MS data acquisition was via selected ion monitoring in negative scanning mode. Quantification was performed using the transitions m/z 333 → 233 and 323 → 221 for armillarisin succinate ester and internal standard, respectively. The method was validated with respect to linearity, accuracy, precision, recovery and stability. This assay has been successfully applied to a pharmacokinetic and tissue distribution study after intravenous injection of ASE in mouse in a dose of 10 mg/kg.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.