Abstract

BackgroundEvaluation of changes in tumor size from images acquired by ultrasound (US), computed tomography (CT) or magnetic resonance imaging (MRI) is a common measure of cancer chemotherapy efficacy. Tumor size measurement based on either the World Health Organization (WHO) criteria or the Response Evaluation Criteria in Solid Tumors (RECIST) is the only imaging biomarker for anti-cancer drug testing presently approved by the United States Food and Drug Administration (FDA). The aim of this paper was to design and test a quality assurance phantom with the capability of monitoring tumor size changes with multiple preclinical imaging scanners (US, CT and MRI) in order to facilitate preclinical anti-cancer drug testing.MethodsThree phantoms (Gammex/UTHSCSA Mark 1, Gammex/UTHSCSA Mark 2 and UTHSCSA multimodality tumor measurement phantom) containing tumor-simulating test objects were designed and constructed. All three phantoms were scanned in US, CT and MRI devices. The size of test objects in the phantoms was measured from the US, CT and MRI images. RECIST, WHO and volume analyses were performed.ResultsThe smaller phantom size, simplified design and better test object CT contrast of the UTHSCSA multimodality tumor measurement phantom allowed scanning of the phantom in preclinical US, CT and MRI scanners compared with only limited preclinical scanning capability of Mark 1 and Mark 2 phantoms. For all imaging modalities, RECIST and WHO errors were reduced for UTHSCSA multimodality tumor measurement phantom (≤1.69 ± 0.33%) compared with both Mark 1 (≤ -7.56 ± 6.52%) and Mark 2 (≤ 5.66 ± 1.41%) phantoms. For the UTHSCSA multimodality tumor measurement phantom, measured tumor volumes were highly correlated with NIST traceable design volumes for US (R2 = 1.000, p < 0.0001), CT (R2 = 0.9999, p < 0.0001) and MRI (R2 = 0.9998, p < 0.0001).ConclusionsThe UTHSCSA multimodality tumor measurement phantom described in this study can potentially be a useful quality assurance tool for verifying radiologic assessment of tumor size change during preclinical anti-cancer therapy testing with multiple imaging modalities.

Highlights

  • Evaluation of changes in tumor size from images acquired by ultrasound (US), computed tomography (CT) or magnetic resonance imaging (MRI) is a common measure of cancer chemotherapy efficacy

  • The World Health Organization (WHO) criteria and the Response Evaluation Criteria in Solid Tumors (RECIST) have been widely used as the only imaging biomarker presently approved by the United States Food and Drug Administration (FDA) for drug testing, the use of functional imaging methods such as Positron

  • The Mark 1 phantom was initially designed to fit into preclinical scanners, it was too large to fit into the bore of preclinical mouse CT and MR units

Read more

Summary

Introduction

Evaluation of changes in tumor size from images acquired by ultrasound (US), computed tomography (CT) or magnetic resonance imaging (MRI) is a common measure of cancer chemotherapy efficacy. Tumor size measurement based on either the World Health Organization (WHO) criteria or the Response Evaluation Criteria in Solid Tumors (RECIST) is the only imaging biomarker for anti-cancer drug testing presently approved by the United States Food and Drug Administration (FDA). The World Health Organization (WHO) criteria and the Response Evaluation Criteria in Solid Tumors (RECIST) have been widely used as the only imaging biomarker presently approved by the United States Food and Drug Administration (FDA) for drug testing, the use of functional imaging methods such as Positron. The verification for radiologic assessment of tumor size change using this QA phantom would allow standardization of imaging protocols prior to animal studies, potentially reducing the number of animals required, increasing study efficiency and decreasing cost

Objectives
Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.