Abstract

Sentinel lymph node biopsy (SLNB) is a staging procedure dependent on accurate mapping of draining lymphatics via tracers. Robot-assisted SLNB enables access to multiple neck levels with a single incision and intraoperative fluorescence guidance to the SLN. Lymphatic mapping in swine was done using a magnetic tracer and fluorescent dye, injected into the tongue. MRI preoperatively mapped lymphatic spread of the magnetic tracer. Dissection was performed using a da Vinci Xi robot guided by fluorescence-imaging of the dye. Robot-assisted SLNB was successfully performed in all animals (n=5). A novel MRI protocol differentiated SLNs (n=6) from lower echelon nodes (n=11) based on flow progression. Fluorescence imaging provided valuable intraoperative guidance and correlated with magnetic-positive nodes. This study demonstrates preclinical feasibility of a robot-assisted approach to SLNB using magnetic and fluorescent tracers in the head and neck, enabling both preoperative mapping and intraoperative guidance.