Abstract

The aim of this study was to determine the best functional position of a transcatheter heart valve (THV) implanted as a valve-in-valve (ViV) procedure in small rapid deployment valves (RDV) in an in vitro model. A 21 mm Perceval, Enable or INTUITY RDV was mounted into a pulse duplicator and a 23 mm balloon-expandable or a self-expanding THV was deployed (valve-in-valve) in two different positions. Under physiological hydrodynamic conditions, the performance of the THV was characterised by mean transvalvular pressure gradient (MPG), effective orifice area (EOA) and regurgitation volume (RV). Leaflet kinematics were assessed with high-speed video recordings, and X-ray images were acquired. All THV/RDV combinations met ISO requirements regarding hydrodynamic performance. In most cases, the higher position of the THV performed better than the lower one in terms of a lower MPG and increased EOA. Leaflet motion of the implanted THV was impaired in the lower position. In contrast, regurgitation volumes were relatively small and similar, regardless of the THV position. ViV implantation of a THV in a small RDV yielded satisfactory hydrodynamic results. In most cases, a high implantation position achieved lower MPG, higher EOA and a reduced risk of impaired THV leaflet function. Fluoroscopy images of the best functional ViV positions are presented as a blueprint for patient procedures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.