Abstract

Relative intensities of gamma rays in the decay of 152Eu, 154Eu and 192Ir were measured precisely with a calibrated Ge(Li) detector. Errors of relative intensities for strong gamma rays are 0.3–0.5%. These multi-gamma-ray sources are useful for the calibration of the Ge(Li) detector. Gamma-ray intensities per decay of 192Ir were derived from the relative intensities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.